Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

  1. Open Motorscope https://microvu.atlassian.net/l/c/s78hqPTr

  2. Initialize the machine if it hasn't been already (“Tools” → “Find Volume”).

  3. Go to “View” → “Testing”.

  4. Open the “Sync Settings” dialog.

  5. Find the check box “XLINENC”, “YLINENC” or “ZLINENC” and enable it for the desired axis.

    Image RemovedImage Added
  6. Close the dialog and click "Apply Sync Settings".

  7. Drive or push the corresponding axis back and forth.

  8. Look at the corresponding coordinate of the Position. It will update each time the index mark is passed.

    Image RemovedImage Added
  9. The position should repeat to the same value when traveling in the same direction. In the opposite direction, it should not deviate by more than 1 count2 counts.

If the position does not repeat in the X or Y axis of a Vertex 261, 341, or 342, replace the encoder.

...