Loosen the tension bolt on the capstan drive.
Finger tighten the tension bolt.
Tighten the tension bolt an additional 1⁄2 – 3⁄4 turn.
Use the pendant to move the axis and check the sync error value using Motorscope Access under Tuning view.
...
a Force gauge to fine tune this process because often the Idler Tension Bolt is tightened passed the 3/4 turn.
For this process:
Start with the machine off and cleaning the band with Acetone.
Finger tighten Idler Tension Bolt.
Tighten Idler Tension Bolt and additional 1/2 turn.
Place a 2.5 L Key in one of the bolts on the Coupler Shaft ( This will Lock the Drive):
Using the Force Gauge push on the locked drive until it slips.
Read the peak force on the gauge. For the X Axis it should be between 50-75 lbs.
If it is under 50 lbs tighten the Idler Tension Bolt, if it is higher than 75 lbs loosen the Idler Tension bolt.
Make sure to move the drive to a different spot after testing with the force gauge this will prevent isolated damage to the band.
Repeat steps d-h as needed.