Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

Number

Item

Micro-Vu Item #

Description

7

Fuse

4A, Rotary stage power input.

8

Fuse

4A, Platen power input.

9

Fuse

1.5A, GPIO1 power input.

10

Fuse

1.5A, GPIO2 power input.

11

Fuse

0.375A, Pendant 5V power input.

12

Fuse

0.375A, Pendant 3.3V power input.

13

LED

N/A

Board firmware state indicators.

Are both LEDs blinking, not necessarily in sync?

  • Yes, indicate the board is operational.
  • No, one LED is not blinking indicates the logic board is defective.
  • Both LEDs are ON and not blinking indicate board is likely not programed or defective.

In Vertex models the Logic Board is located in the back of the machine in the electrical panel:

...

In the Excel models the Logic board is located on the side of the machine in the electrical panel:

...