For Cameras used in the Micro-Vu machine models:

For Excel, 4000, 5000, 6000, 500, 600, 660, 1000,1600,2000,2500

Vertex 110, 120, 130, 210, 220, 230, 310, 320, 330, 410, 420, 430

Cameras should be already programmed to the default settings.  

The only setting you should adjust is white balance. To do so, focus on the light calibration target using high magnification (a piece of white paper will work too if you don't have the lights calibration target handy), select the ringlight, and turn auto lights on. Press " WHITE BAL" once and save the setting. 

White Balance – Onepush (The camera has to be in focus on a white surface, preferably the lights calibration target)

...