White Balance Auto 

Shutter Off 

ACG Manual 

ACG Setting 50% 

Back Light Off 

Iris 60-65% 

Gamma 1.0 

H Reverse Off 

...

Info

For Cameras used in the Micro-Vu machine models:

For Excel, 4000, 5000, 6000, 500, 600, 660, 1000,1600,2000,2500

Vertex 110, 120, 130, 210, 220, 230, 310, 320, 330, 410, 420, 430

Cameras should be already programmed to the default settings.  

The only setting you should adjust is white balance. To do so, focus on the light calibration target using high magnification (a piece of white paper will work too if you don't have the lights calibration target handy), select the ringlight, and turn auto lights on. Press " WHITE BAL" once and save the setting to ONEPUSH.

CCD (JAI) Default Settings

White Balance – Onepush (The camera has to be in focus on a white surface, preferably the lights calibration target)

· Shutter – Off

· AGC – Manual

· AGC setting – Move the cursor to the min then

push the up button 4 times (about 25%)

· Back Light – Off

· Iris – N/A

· Gamma – 1.0

· H Reverse – Off

· Sensitivity mode – Manual (Camera Rev.D&E)

· Sensitive – Off