Info |
---|
The Motor Amplifier board is item #SC8069332 used in Vertex 261, 34X and Excel 51X, 70X, 90X, 130X, 170X, 2500 models. |
...
Micro-Vu Item # | N/A | Z-Axis over current LED: - Green= OK.
- Red= Unconfigured (see status LED #6). Over current detected on axis.
|
2 | LED |
N/A | Y-Axis over current LED: - Green=OK.
- Red= Unconfigured (see status LED #6). Over current detected on axis.
|
3 | LED |
N/A | X-Axis over Current LED: - Green=OK
- Red= Unconfigured (see status LED #6). Over current detected on axis
|
4 | Fuse | 2A, Z-Brake Fuse. |
5 | LED Bank |
N/A | 24V_Main - Illuminated LED indicates the machine is powered.
- Not illuminated
|
indicated the (anything else?)- Could be a defective fuse #7 or defective main connector.
24_MOT - Illuminated LED indicates the motors are powered
|
? - .
- Not illuminated LED indicates failure to configure board or a fault/E-stop has occurred
|
?/& 5V LOCAL - Illuminated LED indicates, the board subsystems are powered and
|
board is booted (ready to operate?)- working correctly.
- Not illuminated LED indicates a device on the board or possibly a device connected to the board is not powered.
|
6 | LEDÂ |
N/AIs Green when - If Status LED is green then the board is properly configured
|
and no faults have occurred does not go green (what are the options? off or green?) after powering up - is off after powering up machine, there is likely an issue with the motor amplifier board.
- If Status LED is red 10 seconds after powering up the machine, there is likely an issue with the amplifier board, logic board, or cable between them.
|
7 | Fuse | Fuse size based on machine model: |
#S81152095 - 81152095 used on Vertex 261,341,342.
- 30A FUSE item
|
#S81152105 - 81152105 used on Excel 51X 70X, 90X, 130X, 170X, 2500.
|
In Vertex models the Motor Amplifier Board is located in the back of the machine in the electrical panel:
...
In the Excel models the Motor Amplifier Board is located on the side of the machine in the electrical panel:
...