This article only applies to Post 2020 Machines
Problem
The optics calibration starts but fails shortly after. InSpec displays an error message saying that the optics calibration failed.
Solution
Open the error.trl file. This file is by default located here: C:\ProgramData\Micro-Vu Corporation\InSpec for Windows
If the error reported is: “No strobe signal”, check the USB cable connecting the camera to the zoom board. It needs to be plugged into CAM1 (J8).
Retry to perform an optics calibration.