Replace bearing blocks on Z-axis
This solution is for Post 2020 Machines
This procedure should ONLY be performed by an authorized Micro-Vu dealer. Please consult your local Micro-Vu authorized sales partner(s) before any operation.
Remove the machine Z cover, shroud, lens, and Z drive.
The Z stage will fall, make sure it is held upward firmly by another operator.
Remove the M4 Screw hard stop at the top of the travel.
There are 8 screws holding the z stage on, 4 in each bearing block. Use 2-4-6 blocks on the glass to support the z stage after it is detached, doing so will avoid the need to undo the igus/cables.
Once the screws are out, the old bearing blocks can be removed from the rail at the top of the travel, and new ones can be installed.
Make sure the reference surfaces on the Z stage that mount to the pads are clean, dry, and free of all dust and particles.
Make sure the reference surfaces on the pads that mount to the Z stage are clean, dry, and free of all dust and particles.
Make sure to press the Z stage against the reference edges of the pads before tightening the four M3 screws that hold each pad in place.
Follow opposite steps to reassemble the machine.
Par-focal
Par-centric
Camera height
Camera squareness
Ringlight alignment
Profile alignment
Optics calibration
Lights calibration
Detailed instructions on these calibrations/adjustments can be found in the Installation Manuals:
Product information is subject to change without prior notice as a result of ongoing technical development