Transfer Settings and Calibrations to A New Installation from An Existing Installation
This article applies to InSpec
Backup Existing Settings and Calibrations
To backup existing settings and calibrations:
Close InSpec.
Navigate to
C:\ProgramData\Micro-Vu\InSpecfolder. This folder might be hidden from Windows Explorer. To navigate to it:Open Windows Explorer.
Paste the path into the address bar. As shown here.
Copy the content of the folder into a secure location.
- If this PC is the same PC used in the previous installation, then follow “Restoring Existing Settings and Calibrations onto The Same PC”
- If this PC is a new PC, then follow “Transfer Existing Settings and Calibrations onto A New PC”
Restoring Existing Settings and Calibrations onto The Same PC
This process only applies if the PC is the same PC used in the previous installation.
Before installing InSpec, have the files in the backed-up folder extracted and ready to use at another location (e.g.
C:\MicroVuProgramDataBackup). Do not create the folderC:\ProgramData\Micro-Vu\InSpecor paste the folder back toC:\ProgramData\Micro-Vu\InSpec.Install the same version of InSpec as before using the installer.
When prompted to “Activate InSpec”, there’s no need to visit the Activation Portal, directly proceed to the next step.
When prompted to “Browse for the license file”, browse to the folder extracted in step 1. and select the file “License.lic”. Proceed with the rest of the installation steps
After the installation is complete, InSpec should be able to be launched.
After InSpec launched without issue, close InSpec immediately.
Copy the rest of the content extracted from the folder in step 1. to
C:\ProgramData\Micro-Vu\InSpecOpen InSpec, settings and calibrations should be restored.
Transfer Existing Settings and Calibrations onto A New PC
To use InSpec on a new PC, the activation process needs to be re-performed on the new PC, which may require a new activation code.
This process will delete settings and calibrations on the target PC.
Install the same version of InSpec as before using the installer, including performing the activation as prompted by the installer.
Close InSpec.
Navigate to
C:\ProgramData\Micro-Vu\InSpecfolder.From the backup location determined above, except the file
License.lic, copy the files and folders then paste into theC:\ProgramData\Micro-Vu\InSpecfolder. When prompted to replace files, click “Yes”.Open InSpec, settings and calibrations should be restored.