The Logic board is item #SC8069331 used in Vertex 261, 34X and Excel 51X, 70X, 90X, 130X, 170X, 2500 models.

Number

Item

Micro-Vu Item #

Description

1

Fuse

6.3A, Logic Board Power Input. 

2

E-Stop Jumper

N/A

Jumper used to bypass in-machine E-Stop.

Only used in Vertex machines.

3

ADIO Enable Switch

N/A

By default should be switched towards ‘Bypass’.

4

ADIO

N/A

N/A

5

LED

N/A

Board power in indicator.

Is it illuminated?

  • Yes, indicates power is delivered to the board.
  • No, indicates power is not being delivered to board. Check Amplifier board is fully powered and operational (SC8069332). Check power cable between logic board and amp board, pins, and connectors.

6

LED

N/A

Board subsystems power indicator.

Are all LEDs illuminated when the machine is powered on?

  • Yes, indicates normal behavior.
  • Not all LEDs illuminated LED #5 not illuminated indicates power not delivered to board.
  • Not all LEDs illuminated and LED #5 is illuminated indicates board and/or connected devices are not operating properly.

Number

Item

Micro-Vu Item #

Description

7

Fuse

4A, Rotary stage power input.

8

Fuse

4A, Platen power input.

9

Fuse

1.5A, GPIO1 power input.

10

Fuse

1.5A, GPIO2 power input.

11

Fuse

0.375A, Pendant 5V power input.

12

Fuse

0.375A, Pendant 3.3V power input.

13

LED

N/A

Board firmware state indicators.

Are both LEDs blinking, not necessarily in sync?

  • Yes, indicate the board is operational.
  • No, one LED is not blinking indicates the logic board is defective.
  • Both LEDs are ON and not blinking indicate board is likely not programed or defective.