Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

  1. Check the lights of the probe module and probe head when attaching and triggering the touchprobe carefully with a finger.

    1. For TP-20:

      1. The PH6 LED should be off when the TP-20 module is not attached

      2. The PH6 LED should be on when the TP-20 module is attached

      3. Triggering the TP-20 module should blink the PH6 probe head LED off, regardless of whether the touchprobe is the active sensor in InSpec or not.

    2. TP-200:

      1. The PH6 LED should be on when the TP-200 module is not attached

      2. The PH6 LED should be on when the TP-200 module is attached

      3. Triggering the TP-200 module when InSpec is not set to touchprobe should:
        the PH6 probe head LED should not blink but it will remain on
        the TP-200 probe module LED should remain off

      4. Triggering the TP-200 module when InSpec is set to touchprobe:
        the PH6 probe head LED should blink off
        the TP-200 probe module LED should light up with a brighten by an amount commensurate with the probe force as it is deflected.

  2. Check the DIP switches of the PI-200 probe interface. Accessing this box might require to disassemble parts of the machine and should only be executed by Micro-Vu dealers.
    The following Image shows the position of the PI-200 on a Vertex machine:

    Image RemovedImage Added


    The following Image shows the position of the PI-200 on an Excel machine:

    Image RemovedImage Added


    When the PI-200 box can be accessed, the DIP switch configuration should be compared to the reference which is presented below.

  3. Additional information and instructions to perform testing with a TP-200 in motorscope can be found at Checking the TP-200 Touch Probe State

    The TP-20 and TP-200 touchprobe installation manual is available at:
    P2 Touch Probe TP20 and TP200 Installation Manual

Filter by label (Content by label)
showLabelsfalse
max5
spacescom.atlassian.confluence.content.render.xhtml.model.resource.identifiers.SpaceResourceIdentifier@141d7
sortmodified
showSpacefalse
reversetrue
typepage
cqllabel in ( "kb-troubleshooting-article" , "post-2020" , "touchprobe" ) and type = "page" and space = "SRV"
labelspost-2020 touchprobe kb-troubleshooting-article

...