  1. Check the lights of the probe module and probe head when attaching and touching the probe carefully with a finger.

    1. For TP-20:

      1. The PH6 LED should be off when the TP-20 module is not attached

      2. The PH6 LED should be on when the TP-20 module is attached

      3. Touching the TP-20 module should blink the PH6 probe head LED off, regardless of whether the touch probe is the active sensor in InSpec or not.

    2. TP-200:

      1. The PH6 LED should be on when the TP-200 module is not attached

      2. The PH6 LED should be on when the TP-200 module is attached

      3. Touching the TP-200 module when InSpec is not set to touch probe should cause:
        the PH6 probe head LED not to blink but it will remain on
        the TP-200 probe module LED should turn on blink

      4. Touching the TP-200 module when InSpec is set to touch probe should cause:
        the PH6 probe head LED to blink off
        the TP-200 probe module LED to brighten by an amount commensurate with the probe force as it is deflected.

  2. Check the DIP switches of the PI-200 probe interface. Accessing this box might require disassembling parts of the machine and should only be executed by Micro-Vu dealers.
    The following image shows the position of the PI-200 on a Vertex machine:


    The following image shows the position of the PI-200 on an Excel machine:


    When the PI-200 box can be accessed, the DIP switch configuration should be compared to the reference which is presented below.

...

  1. Additional information and instructions to perform testing with a TP-200 in motorscope can be found at Checking the TP-200 Touch Probe State

  2. The TP-20 and TP-200 touch probe installation manual is available at:
    P2 Touch Probe TP20 and TP200 Installation Manual

Filter by label (Content by label)
showLabelsfalse
max5
spacescom.atlassian.confluence.content.render.xhtml.model.resource.identifiers.SpaceResourceIdentifier@141d7
sortmodified
showSpacefalse
reversetrue
typepage
cqllabel in ( "kb-troubleshooting-article" , "post-2020" , "touchprobe" ) and type = "page" and space = "SRV"
labelspost-2020 touchprobe kb-troubleshooting-article

...