This article applies only to P2 Machines

  1. Open Motorscope https://microvu.atlassian.net/l/c/s78hqPTr

  2. Initialize the machine if it hasn't been already (“Tools” → “Find Volume”).

  3. Go to “View” → “Testing”.

  4. Open the “Sync Settings” dialog.

    Image Added
  5. Find the check box “XLINENC”, “YLINENC” or “ZLINENC” and enable it for the desired axis.

    Image RemovedImage Added
  6. Close the dialog and click "Apply Sync Settings".

  7. Drive or push the corresponding axis back and forth.

  8. Look at the corresponding coordinate of the Position. It will update each time the index mark is passed.

    Image Removed

...


  1. The position should repeat to the same value when traveling in the same direction. In the opposite direction, it should not deviate by more than 2 counts.

...