This article only applies to P2 Machines

Problem

...

The optics calibration starts but fails shortly after. InSpec displays an error message saying that the optics calibration failed.

...

Solution

  1. Open the error.trl file. This file is by default located here: C:\ProgramData\Micro-Vu Corporation\InSpec for Windows

  2. If the error reported is: “No strobe signal”, check the cable connecting the camera to the zoom board. It needs to be plugged into CAM1 (J8).

    Image Added