This article applies only to P2 Post 2020 Machines Excel (X,Y and Z-axis) and Vertex (Z-axis only)

Procedure

Encoder Alignment

  1. Remove all covers to be able to access the encoder readhead of the corresponding axis.

  2. Make sure that the calibration LED on the encoder readhead is not illuminated. If it is illuminated, press and hold the calibration button on the interface for >3s, or until the calibration LED is no longer illuminated.

    Image Modified

  3. Make sure the center LED of the encoder readhead is illuminated green. If it is not illuminated green, the yaw and/or pitch of the readhead have to be readjusted. If that is the case, loosen the two M4 screws holding the readhead bracket carefully and adjust the readhead until the center LED indicates green.

    Image Modified

  4. Watch the colored light on the encoder interface/connector (located within the electronics enclosure of the machine) while driving/pushing the corresponding axis across the entire travel.

    Image Modified



    The color indicates the acceptability of the readhead distance relative to the scale:

    Image Modified

  5. Adjust the readhead distance in case the light indicates a distance that is not acceptable,
    by using the fine adjustment screws of the encoder readhead bracket.

...

  1. Image Added
Note

It is considered failing if the light changes to red or orange at any point during the travel!

Info

Additional information on how to align the encoder readhead can be found here: https://microvu.atlassian.net/l/c/GREVRkuV

Encoder Calibration

  1. Press the calibration button on the encoder interface/connector for less than 2 seconds, using i.e. a 2mm allen key. The calibration LED on the readhead will single flash yellow.

...

  1. Image AddedImage Modified

    https://resources.renishaw.com/en/details/installation-guide-tonic-t103x-rtlc20-fastrack-linear-encoder-system--111511

  2. Calibrate the encoder without crossing an index mark (incremental signal calibration).
    Slowly drive the corresponding axis back and forth, using the pendant or pushing the axis, ensuring that you stay within the limit marks until calibration LED begins to double flash yellow.

  3. Phase the encoder by driving/pushing the axis across the reference mark, until the calibration light stops double flashing.

Info

If the calibration LED does not stop double flashing, press the calibration button to manually exit the calibration routine and power down the machine. Once the system has been powered down, press and hold the encoder calibration button whilst turning the system on. The calibration led will flash several times indicated factory defaults are restored. Repeat the calibration steps.

...

Filter by label (Content by label)
page
showLabelsfalse
max5
spacescom.atlassian.confluence.content.render.xhtml.model.resource.identifiers.SpaceResourceIdentifier@141d7
showSpacefalse
sortmodified
showSpacetypefalsepage
reversetruetype
labelskb-troubleshooting-article
cqllabel = "encoder" and space = "SRV"labelskb-troubleshooting-article
Page Properties
hiddentrue

Related issues