  1. Open Motorscope (Motorscope Access) and go to the testing page by clicking on View > Testing.

  2. Open the Synchronization Settings and enable the CAM1_OUT1 (Rising) sync signal.

  3. Set the LB → ZB Channel Output Trigger Mode to Periodic at 28Hz.

  4. Enable CAM1_OUT1 on Channel 1.

  5. Click Okay then click Apply Sync Settings. You should see some information counting up, indicating the triggering is active.

Untitled-20240312-150940.png

  • If the trigger values

  1. Leave MotorScope open. Open IC Capture and connect to the machine’s camera.

  2. Set the correct settings in the Device Properties:

    1. Enable Trigger

    2. Enable IMX Low-Latency Mode

...

  1. The camera should be streaming at 28 FPS. Verify the camera is live in IC Capture by e.g. shining a light into the lens.

...