No Camera Image in InSpec
This article is for Post 2020 Machines
Problem
InSpec displays no camera image.
InSpec Classic displays the camera image, without issues.
Troubleshooting A
Attempt to reconnect the machine by going to Machine > Machine Library, selecting the connected machine, and clicking Connect.
- If this does not resolve the problem proceed to Troubleshooting B.
Troubleshooting B
Open Motorscope (Motorscope Access) and go to the testing page by clicking on View > Testing.
Open the Synchronization Settings and enable the CAM1_OUT1 (Rising) sync signal.
Set the LB → ZB Channel Output Trigger Mode to Periodic at 28Hz.
Enable CAM1_OUT1 on Channel 1.
Click Okay then click Apply Sync Settings. You should see some information counting up, indicating the triggering is active.
- When the signal is received properly, the camera should stream frames at the rate configured in MotorScope.
Leave MotorScope open. Open IC Capture and connect to the machine’s camera.
Apply the correct settings in the Device Properties:
Enable Trigger
Enable IMX Low-Latency Mode
Set the correct format along the top:
YUY2, 2X Skipping for color camera systems.
Y800, 2X Binning for monochrome camera systems as displayed in Image ‘Step 8’.
FPS to 232.06
The camera should be streaming at 28 FPS. Verify the camera is live in IC Capture by e.g. shining a light into the lens.
- If the FPS can not be set to 28 then proceed to change the trigger cable and camera. Find the item numbers here: https://microvu.atlassian.net/l/cp/kkuTtwmb
Product information is subject to change without prior notice as a result of ongoing technical development