Info |
---|
This troubleshooting article applies only to post Post 2020 machines Machines |
If the machine is not operating and the” information light” next to the power button at the rear status panel of the machine and the firmware LED on the logic board flashes with the following eight-bit pattern as described here:
...
The following items provide step by step instructions on how to break down the cause of the issue:
...
Make sure the pendant
...
is plugged in
...
when the
...
machine is booting up
If the pendant was not unplugged subsequently, there is a hardware issue:
Pendant, Pendant Cable or Pendant Board are defective
Logic Board is defective
Note |
---|
plugging in the pendant while the machine is powered up (hot-plugging) is not supported! |
For more details and how to perform debugging, see the following article:
https://microvu.atlassian.net/wiki/x/KQAhug
Related articles
Filter by label (Content by label) | |||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Page Properties | ||
---|---|---|
| ||
|