Info |
---|
This article applies to the following machine models: Excel 50X, 60X, 66X, 10XX, 16XX, 2000, 2500 Vertex 251, 31X SOL 161, 31X |
A machine calibration is not necessary after exchanging the logic board.
Before replacing the board, create a backup of the Micro-Vu machine information as a precaution. To do this, save the folder named "InSpec for Windows" in an external memory. This folder is located here: C:\ProgramData\Micro-Vu Corporation
Turn the machine off, then proceed to disconnect all the cables from the logic board.
...