Info |
---|
This is These instructions are for Pre 2020 Machines only and Post 2020 Machines |
InSpec Classic software has the ability to “Home” the stage by pressing the F8 button on the keyboard. When pressing F8 on the keyboard the machine moves to a loading position, which varies between Vertex and Excel systems.
The load position can be changed to a different camera/stage position following these steps:
Open the Machine.ini file located by default in the following address: C:\ProgramData\Micro-Vu Corporation\InSpec for Windows\”MACHINE SERIAL NUMBER”
Add to the Machine.ini file:
on Pre 2020 Machines :
[Alpha]
Load Position X=Off
Load Position Y=Min
Load Position Z=Max
on Post 2020 Machines :
[PH2]
Load Position X=Off
Load Position Y=Min
Load Position Z=Max
These are the values that can be set for each axis to control the Load Positionload position:
Off - Do not move that specific axis
...