Skip to end of metadata
Go to start of metadata

You are viewing an old version of this page. View the current version.

Compare with Current View Page History

« Previous Version 9 Next »

Post 2020 machines signal error conditions with audible and/or visual codes which are explained in the table below.

Error conditions are signaled by blinking of the “information light” next to the power button of the machine and the firmware LED on the logic board. High priority codes will be accompanied by audible beeping of the same pattern.

Operation Modes

Two general operation modes can be differentiated:

  • Normal operation is indicated by:
    1/2 second on, followed by
    1/2 second off, continuous blinking.
  • Error condition is indicated by:
    2 second off, followed by
    1/4 second (0, short) or 1 second (1, long)

Fault Codes

The pulses in error condition indicate an eight-bit binary error-code with:

  • short pulse = 0 which is 1/4 second long
  • long pulse = 1 which is 1 second long

Low Priority Codes

Visual signal of the information light and firmware LED only

Code

Meaning

0000 0001

An emergency stop button has been depressed. This is not strictly an error condition.                             

0000 0010

The pendant is missing or malfunctioning.                                                                         

0000 0100

The machine is in standby mode.                                                                                   

0000 0101

The x-axis has encountered an encoder limit.                                                                      

0000 0110

The y-axis has encountered an encoder limit.                                                                      

0000 0111

The z-axis has encountered an encoder limit.                                                                      

0000 1000

Stall, x-axis. The motor could not move the stage.                                                                

0000 1001

Stall, y-axis.                                                                                                     

0000 1010

Stall, z-axis.                                                                                                    

0000 1100

Following error fault, x-axis. The following error for the motion control system has exceeded the fault threshold.

0000 1101

Following error fault, y-axis.                                                                                    

0000 1110

Following error fault, z-axis.                                                                                    

0000 1111

Synchronization error fault, x-axis. The linear and rotary encoder differ too much.                               

0001 0000

Synchronization error fault, y-axis.                                                                              

0001 0001

Synchronization error fault, z-axis.                                                                              

0001 0011

Overcurrent, x-axis. The amplifier board has reported that the axis current has exceeded its configured maximum.  

0001 0100

Overcurrent, y-axis.                                                                                              

0001 0101

Overcurrent, z-axis.                                                                                              

0001 0110

Servo overtime fault. The regulator failed to be run in a timely manner.                                          

0001 1000

Find volume fault. The find volume process for the x, y, and z-axis failed.                                       

0001 1001

Find range fault. The find range process for the zoom motor failed.                                               

0001 1010

Platen device fault. The platen has encountered an error and is in a faulted state.                               

0001 1011

Rotary device fault. The rotary has encountered an error and is in a faulted state.                               

High Priority Codes

Visual signal of the information light and firmware LED in combination with accoustic signal of the pieze speaker.

High priority codes override low priority codes

Code

Meaning

0010 0000

Encoder failure, x-axis.                                           

0010 0001

Encoder failure, y-axis.                                           

0010 0010

Encoder failure, z-axis.                                           

0010 0100

Encoder reversed, x-axis.                                          

0010 0101

Encoder reversed, y-axis.                                          

0010 0110

Encoder reversed, z-axis.                                          

0010 1100

Runaway, x-axis. The stage moved unexpectedly.                     

0010 1101

Runaway, y-axis.                                                   

0010 1110

Runaway, z-axis.                                                   

0011 0000

Fan failure, enclosure.                                            

0011 0001

Amplifier board configuration failure.                             

0011 0010

Define, Tuning, or Accessory file is successfully read but invalid.

0011 0011

A platen was detected, but communication to it was interrupted.    

  • No labels