Camera Debugging Instructions
These steps help to diagnose and collect helpful information when encountering camera issues while using a Micro-Vu machine.
This article applies to Post 2020 Machines
This procedure should only be performed by authorized Micro-Vu dealers
Instructions
Follow the instructions below to bypass the breakout board and connect the camera straight to the logic board.
Power off machine
Remove the Z-axis cover
Unplug the USB connector 1A from the zoom breakout board
Unplug the USB connector 2A from the camera
Plug USB connector 1A into the camera
Power on Machine
Download IC Capture: https://www.theimagingsource.com/en-us/support/download/iccapture-2.5.1557.4007/
Open InSpec or ICcapture to test camera connection