Skip to end of metadata
Go to start of metadata

You are viewing an old version of this page. View the current version.

Compare with Current View Page History

Version 1 Next »

Problem

This article applies to P2 machines with the additional VAS

The VAS Sensor does not show up in InSpec, is detected but does not measure or the calibration fails

Solution

The following steps describe how to check the cable that connects the VAS to the machine.

  1. Power down and unplug the machine

  2. Remove the Z-axis cover of the machine

  3. remove the Zoom Board cover

  4. Identify the cable that connects the VAS to the Zoom Board

  5. Doublecheck the cable combination and color-coding of the connectors
    Pay special attention to the white and green cables

In case cables are connected incorrectly, the cables can be swapped. For more information contact the support.

Another potential issue could be a cable inserted too far into the green connector like shown below

If this might be the case, press down the orange lever with a slot screw driver, remove and carefully reinsert the cable into the connector.

Power on the machine, launch InSpec and test if the VAS measurement values are displayed.

Filter by label

There are no items with the selected labels at this time.

  • No labels