Skip to end of metadata
Go to start of metadata

You are viewing an old version of this page. View the current version.

Compare with Current View Page History

Version 1 Next »

This article applies only to Post 2020 Machines Vertex X- and Y-Axis

Procedure

An orange spacer is required to perform the alignment procedure which can be requested here.

Encoder Alignment

  1. Remove all covers to be able to access the encoder readhead of the corresponding axis.

  2. Make sure that the stage and the encoder readhead is positioned at the center of its travel.

  3. Loosen the two screws that hold the readhead in place

  4. Place the orange spacer between readhead and encoder scale as shown in the image below:

  5. Push readhead up against the spacer and the encoder scale and tighten the two readhead screws

  6. Remove the spacer, move stage to the left end of the travel and check gap by placing the spacer between readhead and scale.

  7. Loosen the screw that holds the encoder scale on the corresponding side of the travel in place.

  8. Push the encoder scale down against the spacer and the readhead while tightening the screw again
    A T-handle can be used to push the scale down by sliging it in the hole on the side of the scale like shown in the image below:

  9. Repeat steps 6, 7 and 8 on the other end of the travel.

  • No labels