This article is for Post 2020 Machines
Problem
InSpec displays no camera image.
InSpec Classic displays the camera image, without issues.
Troubleshooting A
Attempt to reconnect the machine by going to Machine > Machine Library, selecting the connected machine, and clicking Connect.
- If this does not resolve the problem proceed to Troubleshooting B.
Troubleshooting B
Open Motorscope (Motorscope Access) and go to the testing page by clicking on View > Testing.
Open the Synchronization Settings and enable the CAM1_OUT1 (Rising) sync signal.
Set the LB → ZB Channel Output Trigger Mode to Periodic at 28Hz.
Enable CAM1_OUT1 on Channel 1.
Click Okay then click Apply Sync Settings. You should see some information counting up, indicating the triggering is active.
- If the trigger values
Leave MotorScope open. Open IC Capture and connect to the machine’s camera.
Set the correct settings in the Device Properties:
Enable Trigger
Enable IMX Low-Latency Mode
Set the correct format along the top:
YUY2 with 2X skipping for color (Image ‘Step 8’ shows settings for a monochrome camera)
FPS to 232.06
The camera should be streaming at 28 FPS. Verify the camera is live in IC Capture by e.g. shining a light into the lens.
- If the FPS can not be set to 28 then proceed to change the trigger cable and camera. Find the item numbers here: https://microvu.atlassian.net/l/cp/kkuTtwmb