  1. Open Motorscope (Motorscope Access) and go to the testing page by clicking on View > Testing.

  2. Open the Synchronization Settings and enable the CAM1_OUT1 (Rising) sync signal.

  3. Set the LB → ZB Channel Output Trigger Mode to Periodic at 28Hz.

  4. Enable CAM1_OUT1 on Channel 1.

  5. Click Okay then click Apply Sync Settings. You should see some information counting up, indicating the triggering is active.

Untitled-20240312-150940.png

  • If the trigger values When the signal is received properly, the camera should stream frames at the rate configured in MotorScope.
  1. Leave MotorScope open. Open IC Capture and connect to the machine’s camera.

  2. Set Apply the correct settings in the Device Properties:

    1. Enable Trigger

    2. Enable IMX Low-Latency Mode

...

  1. Set the correct format along the top:

    1. YUY2 with , 2X skipping Skipping for color (camera systems.

    2. Y800, 2X Binning for monochrome camera systems as displayed in Image ‘Step 8’ shows settings for a monochrome camera).

    3. FPS to 232.06

...