These instructions apply to P2 Machines

When replacing the Logic board on a P2 Machines please note:

When purchasing a spare logic board (item SC8069331) it will include a jumper inside the bag. This jumper should be plugged into the logic board if the board is installed on a P2 Machines Vertex model only.

If the logic board is installed on an Excel machine, the jumper should not be installed on the logic board.

After replacing the logic board on the machine follow these steps to set up the logic board in Motorscope Access

  1. After replacing the Logic Board turn the machine on and open Motorscope.

  2. Motorscope should open in the Configuration view. Else click on View-Configuration:

  3. Select the machine model from the drop-down menu.

  4. Type the machine serial number and enter the information that has not been automatically configured such as: ENTER INFO HERE

  5. Select Temporarily Configure.

If the machine works well after replacing the logic board and configuring the machine you should make the configuration permanent by clicking on “Permanently Configure”.