These instructions are for Pre 2020 Machines

InSpec software has the ability to “Home” the stage by pressing the F8 button on the keyboard. When pressing F8 on the keyboard the machine moves to a loading position, which varies between Vertex and Excel systems.

The load position can be changed to a different camera/stage position following these steps:

  1. Open the Machine.ini file located by default in the following address: C:\ProgramData\Micro-Vu Corporation\InSpec for Windows\”MACHINE SERIAL NUMBER”

  2. Add to the Machine.ini file:

[Alpha]

Load Position X=Off

Load Position Y=Min

Load Position Z=Max

These are the values that can be set for each axis to control the load position:

Off - Do not move that specific axis

Min – At the negative limit (close to it)

Mid – Halfway between the limits of the machine

Max – At the positive limit (close to it)

### - Distance from the machine coordinate system in mm