Skip to end of metadata
Go to start of metadata

You are viewing an old version of this page. View the current version.

Compare with Current View Page History

« Previous Version 3 Next »

Problem

This article includes instructions to troubleshoot the touchprobe on P2 Machines featuring a PH6 probe head and a TP-20 or TP-200 probe module.

The following picture shows the LED on the PH6 probe head:

PICTURES OF TO-20 AND TP-200 PROBE MODULES SHOULD BE ADDED HERE, SHOWING THE LED

Solution

Perform the following step in case the touchprobe doesn’t trigger or if the device exhibits unintended behavior.

  1. Check the lights of the probe module and probe head when attaching and touching the probe carefully with a finger.

    1. For TP-20:

      1. The PH6 LED should be off when the TP-20 module is not attached

      2. The PH6 LED should be on when the TP-20 module is attached

      3. Touching the TP-20 module should blink the PH6 probe head LED off, regardless of whether the touchprobe is the active sensor in InSpec or not.

    2. TP-200:

      1. The PH6 LED should be on when the TP-200 module is not attached

      2. The PH6 LED should be on when the TP-200 module is attached

      3. Touching the TP-200 module when InSpec is not set to touchprobe should cause:
        the PH6 probe head LED not to blink but it will remain on
        the TP-200 probe module LED should turn on

      4. Touching the TP-200 module when InSpec is set to touchprobe should cause:
        the PH6 probe head LED should blink off
        the TP-200 probe module LED should brighten by an amount commensurate with the probe force as it is deflected.

  2. Check the DIP switches of the PI-200 probe interface. Accessing this box might require to disassemble parts of the machine and should only be executed by Micro-Vu dealers.
    The following Image shows the position of the PI-200 on a Vertex machine:


    The following Image shows the position of the PI-200 on an Excel machine:


    When the PI-200 box can be accessed, the DIP switch configuration should be compared to the reference which is presented below.

  3. Additional information and instructions to perform testing with a TP-200 in motorscope can be found at Checking the TP-200 Touch Probe State

    The TP-20 and TP-200 touchprobe installation manual is available at:
    P2 Touch Probe TP20 and TP200 Installation Manual

  • No labels