Replace a Zoom Lens
This procedure should be done by an authorized Micro-Vu dealer
Before unplugging any cables or components from the machine make sure the machine is turned off.
Move the Z axis all the way to the bottom, as close to the glass as possible.
Mark with a pencil or piece of tape, where the Z-stage end aligns with the Optics Shroud, so this can be mounted at the same height when assembling it back together.
Follow the steps shown in the video attached to remove the zoom assembly from the machine.
After replacing the Zoom assembly, please perform the following procedures as described in the machine installation and maintenance guide:
Par-focal Adjustment
Par-centric Adjustment
Camera Height
Camera Squareness
Optics Calibration
Lights Calibration
Stage verification
Stage Calibration if the results of the stage verification are not satisfactory
Product information is subject to change without prior notice as a result of ongoing technical development