Configuration File Invalid Error
Problem
This troubleshooting article applies only to Post 2020 Machines
If the machine is not operating and the” information light” at the rear status panel of the machine and the firmware LED on the logic board flashes with the following eight-bit pattern as described here:
0011 0010
Where 0 indicates a short pulse which is 1/4 second and 1 indicates a long pulse which is 1 second long.
Since this is a high priority error, the internal piezo speaker will also output an acoustic signal with the same pattern.
This indicates that one or more of the configuration files on the Logic Board are corrupted.
Solution
Check the following hardware:
Replace the Logic Board
Related articles
Product information is subject to change without prior notice as a result of ongoing technical development