Configuration File Invalid Error

Problem

This troubleshooting article applies only to Post 2020 Machines

If the machine is not operating and the” information light” at the rear status panel of the machine and the firmware LED on the logic board flashes with the following eight-bit pattern as described here:

0011 0010

Where 0 indicates a short pulse which is 1/4 second and 1 indicates a long pulse which is 1 second long.

Since this is a high priority error, the internal piezo speaker will also output an acoustic signal with the same pattern.

 

This indicates that one or more of the configuration files on the Logic Board are corrupted.

Solution

Check the following hardware:

  • Replace the Logic Board

Related articles

Product information is subject to change without prior notice as a result of ongoing technical development