Find Range Error

Problem

This troubleshooting article applies only to Post 2020 Machines

If the machine is not operating and the” information light” at the rear status panel of the machine and the firmware LED on the logic board flashes with the following eight-bit pattern as described here:

0001 1001

Where 0 indicates a short pulse which is 1/4 second and 1 indicates a long pulse which is 1 second long.

This error might occur during initialisation if the machine fails to complete the range finding routine of the zoom lens.

Solution

The following links provide a guide on how to debug the zoom lens:

Zoom Initialization Failure/ Zoom Lens Stall

https://microvu.atlassian.net/wiki/spaces/SRV/pages/1629782077

Related articles

Product information is subject to change without prior notice as a result of ongoing technical development